EE 420L
Analog Integrated Circuit Design Laboratory
Laboratory Report 2: Operation of a Compensated Scope Probe

 

AUTHOR: Bryan Kerstetter

EMAIL: kerstett@unlv.nevada.edu

FEBRUARY 6, 2019


General Overview

Probing can either make or break an electronics project. Poor probing can lead one to believe that their project is not working. This laboratory educates students about good probing habits while introducing the compensated scope probe.


Prelab

I watched Dr. Baker’s video about the compensated scope probe while reviewing the concepts covered on my Laboratory Report 1.


Description of Laboratory Procedures

Waveforms of a 10:1 Compensated Scope: Undercompensated, Compensated, and Overcompensated

For the entirety of this laboratory, the Tektronix TPP0200 Voltage Probe (Figure 3) was used. This particular probe is classified as a 10:1 compensated scope probe. A screw driver can be used to adjust the scope capacitance. I assume, this is done by adjusting the distance of charged plates. Or possibly, adjusting the overlap area of the two plates. These concept can be explained in Equation 1.

               [1]

Variable

Meaning

Units

A

The Overlap Area of the Two Plates

Separation Distance

Permittivity of Free Space

C

Capacitance

Farads

Figure 1

Figure 2

 IMG_2406

Figure 3

LTspice Model

Here we can see that the capacitance of the scope probe (C1), determines whether the scope probe is compensated properly. If C1 is to large then the scope probe becomes overcompensated. If C1 is too small, then the scope probe is undercompensated. Figure 4 shows three cases of scope probes where the respective probes are compensated, undercompensated, and overcompensated. Figure 5 shows the inspected traces, where we can see the actual input voltage vs the inspected voltage traces. Note that the inspected voltages are 1/10 the input voltage. Hence, as to why the scope probe in theses simulations are classified as a 10:1 compensated scope probe. Figure 6 zooms on the compensated, undercompensated, and overcompensated voltage traces. Experimentally, we attached three separate compensated scope probes to the oscilloscope test signal. Then we adjusted the compensation of each probe (see Figure 2) such that there were compensated, undercompensated, and overcompensated scope probes all to be observed. Our experimental results can be seen in Figure 7.

Figure 4

Figure 5

Figure 6

Experimental Results

lab2_correctUNDERover

Figure 7

Attenuation Adjustment on the Tektronix TDS 2014 Oscilloscope

The probe type can be adjusted on the Tektronix TDS 2014 Oscillscope by the following procedure demonstrated in the list below and in Figure 8.

1.     Push the CH 1 MENU button

2.     Push the buttoned demarked by BOX 2 in Figure 8 to toggle probe adjustments

3.     Look at the probe adjustment (BOX 3 in Figure 8)

Figure 8

Draft of 10:1 Scope Probe Schematic

The Tektronix TPP0200 Voltage Probe datasheet, the capacitance ranges of the scope probe tip can be seen in Figure 9.

Figure 9

The oscilloscope used in this lab (from this point onwards) is the Tektronix TDS 2014C. According to oscilloscope datasheet (figure 10), the input impedance of the oscilloscope is a 1 MΩ resistor in parallel with a 20 pF capacitor.

Figure 10

Additionally, we measured the capacitance of the cable to be 82 pF. We then compensated the scope probe properly. To determine reasonable/correct values, I used the equation derived in my first laboratory report (Equation 15).

          [2]

          [3]

Therefore, we can say that a reasonable scope probe capacitance is 11.33 pF for our 10:1 scope probe. Figure 11 shows our drafted schematic with reasonable/correct values.

Figure 11

10:1 Scope Probe Attenuation Factor Proof

We may then simplify Figure 11 down to what is seen in Figure 12.

Figure 12

                   [4]

Based upon Figure 12, we can see that .

                   [5]

                   [6]    

                           [7]

Also, based upon Figure 12, we can see that  and .

          [8]    

 

Therefore, the attenuation factor of our scope probe has been proved.

Device an Experiment to Measure the Capacitance of a Coaxial Cable Without a Farad Meter

Theory

By this point, it has been demonstrated that a coaxial cable has capacitance. This capacitance can be measured with a farad meter (multimeter, LCR meter, etc.). However, how might one measure capacitance without a farad meter? Previously, in Laboratory 1, we covered a basic review of RC circuits. This knowledge will allow us to measure capacitance with just an oscilloscope. Equations 9-12 and Figure 13 demonstrate the knowledge required for this experiment.

                    [9]

                  [10]

            [11]

          [12]

Figure 13

Calculating Expected Values

With a farad meter (LCR Meter), we measured the capacitance of the cable to be 173 pF.

R1 = 100 kΩ                            [13]

Ccable = 173 pF                      [14]

           [15]

           [16]

              [17]

           [18]

In preparation for our experiment, we may determine what the frequency of our Vpulse input square wave.

                        [19]

We also decide that pulse should be from 0 V to 5 V.

Figure 14

In the experiment we connected the circuit as seen in Figure 14. The oscilloscope output can be seen in Figure 17. A 1.8-meter cable was used.

IMG_2395

Figure 15

IMG_2396

Figure 16

IMG_2394

Figure 17

Comparing Hand Calculations and Experimental Results

 

Hand Calculations

Experimental Results

Figure 18

Experimental Calculation of Cable Capacitance

To calculate the cable capacitance without the use of a farad meter we may use any one of Equations 15-18. In this laboratory we decided to use Equation 17.

              [18]

                            [19]

Comparing Measured and Calculated Cable Capacitances

 

Measured Using LCR Meter

Calculated From Trace Seen On Oscilloscope

Cable Capacitance

173 pF

Figure 19

Cable Length and Capacitance

In the pre-laboratory lecture, Dr. Baker explained that a good estimate for the capacitance of a cable is 100 pF/m. In our experiment we used a 1.8 meter (~6 feet) long coaxial cable. According, to the estimate (100 pF/m), our cable should have a capacitance of ~180 pF.  Our, measured and calculated results are in agreement with the estimate of 100 pF/m (See Figure 19). 

Probing a Voltage Divider with a Compensated Scope Probe and a Coaxial Cable

Figure 20

We then constructed a voltage divider as described in Figure 20. We then probed the circuit at the ProbingPoint with both a compensated oscilloscope probe and merely a coaxial cable.

Compensated Scope Probe

IMG_2398

Figure 21 Voltage divider probed with a compensated scope probe.

IMG_2399

Figure 22: Waveform of the voltage divider probed with a compensated scope probe.

Probing the voltage divider with a compensated scope probe compensates for the large cable capacitance and allows for a smaller RC time constant. Thereby, we can see the output of the voltage divider beginning to charge the capacitance and partially discharging. Figure 23 and 24 show this case in LTspice.

Figure 23

Figure 24

Coaxial Cable

IMG_2400

Figure 25: Voltage divider probed with a coaxial cable.

IMG_2401

Figure 26: Waveform of the voltage divider probed with a coaxial cable.

Using a normal coaxial cable, a large capacitance is applied to the test point. This large capacitance coupled with the impedance of the oscilloscope, leads to a large RC time constant. A frequency of 1 MHz is so large that, the signal is never able to ever charge fully. Thereby, giving us what appears to be a flat voltage trace. Figure 27 and 28 shows the case of probing the voltage divider with merely a coaxial cable in LTspice.

Figure 27

Figure 28

Probing a PCB Test Point: Multi-Signal & Multi-Attenuation (MSMA) PCB Probing System

The MSMA PCB Probing System is how I would implement a test point on a PCB so that a known length of cable could be could be connected directly to the board without loading the circuitry on the board. Figure 29 explains the probing system. We have n number of onboard test signals. These test signals would be connected to ports, in which standard compensated oscilloscope probes could be connected to. However, this system also allows the ability to connect ordinary cables to the board without loading the circuitry on the board. A switch or a jumper cable would be used to select the test signal to be observed. Then another switch or jumper would be used to select the desired attenuation factor (created by n pairs of resistor-capacitor cascodes). The capacitor contained within each resistor-capacitor cascode will be a variable capacitor within a certain range. The variable capacitors would then be adjusted based upon the length of the cable connected to the PCB Cable Terminal.

Figure 29


 

Return

Return to EE 420L Students

Return to EE 420L Spring 2019 Page

Return to Dr. Baker’s CMOSedu homepage