Lab 2 - EE 420L
Dwayne K. Thomas
kendaleman@gmail.com
2/5/2015
Operation of a Compensated Scope Probe
The Probes used in these labs, have an inherent capacitance due to the
Probe tip as well as the cable attaching the probe to the scope.
Because of this, we need to calibrate our probes before we hook
it up to our circuit so that we get our correct readings. We
use the 5V square wave calibration provided on the scope to accomplish
this. The calibration screw we turn with a small (preferably plastic)
flathead screw driver is either located at the probe head, or it can be
located at the end of the cable with the coaxial conncection that hooks
to our scope. Below is an example of an Overcompensated,
undercompensated, and a correctly compensated probe.
But before we calibrate the probe, we must tell the scope if our probe
is set to 1X or 10X. We will have to do this in the individual
Channel menu our probe is connected to. The scopes we use in this
lab do not automatically sense the 1X or 10X.
| This is the scope reading of an Overcompensated probe |
| This is the scope reading of an Undercompensated probe |
| This is the scope reading of a Correctly Compensated probe |
The 10:1 ratio for a probe
The following schematic and calculations show how we get a 10:1 ratio
for Vout/Vin when we have 9Megohm resistor in parallel with an about
12pF capacitor for the probe tip. This combined impedance which
we will call Z1 is the impedance of the probe tip. The cable
along with the connector to the scope has a combined capacitance of
about 110pF which is in parallel with the input resistance of the scope
which is 1Megohm. We will call this impedance Z2. The
calculations show that Vout/Vin ~ 0.1
Capacitance of a 3-foot wire Experiment
The
following is an experiment performed in order to determine the
capacitance of a 3-foot wire by measuring the phase shift displayed on
our scope.
| As we can see, the phase shift measured with our scope shows us a 36.7 degree shift.
|
| When we measured the capacitance of the 3-foot wire, we received 131pF. This is close to our calculated Capacitance of 114pF. |
| By wiring the 3-foot cable in parallel with our probe
tip, and all in series with a 100kohm resistor, our total Capacitance is the sum of the 110pF of the probe, and
the cable. |
Voltage Divider with 100Kohm Resistors
In
this experiment, we show that uncompensated probes can significantly
affect the measurement results of our circuit due to the loading
effects of our measuring device. We accomplish this by building a
voltage divider and measuring first with just a 3-foot cable and then
with a compensated probe.
| |
The output voltage from the Uncompensated probe shows our pk-pk voltage at 38mV | The
output voltage from the compensated probe shows our pk-pk voltage
at 260mV. A much a higher reading from the uncompensated probing. |
Return to EE420 labs