Lab 2 - ECE 421L 

Authored by Nicholas Banas,

Banasn1@unlv.nevada.edu

2/5/15 

  

This lab is an introduction to the construction of a standard compensated voltage probe.

Correct Probe Compensation

 Due to inherent inherent non-uniformities in the materials used to construct compensated voltage probes, they must be manually adjusted, or compensated, for each piece of test equipment they are connected to.  This is done with an adjustment on either the probe tip or connector.  For the Tectronix scopes used in this lab, the probe type is set in each channel's menu.  Once the type is set , the probe can be compensated using the probe test function and using the adjustment to obtain a correct reading.  Here are a few examples of what a compensation reading looks like:
 
An under-compensated probe:A correctly compensated probe:An over-compensated probe:

 To understand what is happening with compensation, it is necessary to know what the basic construction of a compensated probe looks like.  Below is an example of a basic 10:1 probe schematic:

 

 
Showing the AC Analysis for the V at the probe tip and the V seen by the scope.
 

 

 
Vprobe-tip to Vscope-in is 1:0.1 or 10:1. It is worth noting that at 1 VAC the current the circuit being tested must supply to the scope is 0.1 uA + j67.86 uA, which may significantly impact the circuits performance.

Cable Capacitance

Coaxial cables have a built in capacitance due to their construction that is proportional to their length.  This can be measured with a capacitance meter, or we can use a test setup with the scope and compensated probe.  The pictures below show the readings from each method for the same cable.

 

Scope measurement Meter measurement
 

  
 The scope measurement calculates the capacitance by tan(ɸ) = 1/RjωC => C = 1/Rjωtan(ɸ).  For this cable the scope calculated .213nF vs .131nF for the meter.  While these measurements are off, it may be due to connection effects from the different setups.

Probe effects on measurements

The two plots below are measurements of the same circuit taken with a strait cable and a compensated probe.  There is a large difference, ~6.5x in the output (ch3 vs ch2), due to the load each measurement device put on the circuit.  This is one of the great benefits to using the compensated probe.  Ideally a 10x probe will have 10x the impedance as a strait cable.  Real world factors reduce this (to ~6.5x in this case) but the improvement is still significant.  The impedance can be improved further using larger scale probes if needed (ie. 100x).
 
Measurement with a strait cableMeasurement with a compensated probe

 

Circuit Board Test Point

Finally, a test point may be instituted on a PCB so that a compensated probe is not needed to test the circuit.  Essentially an interface is created so that testing the circuit has no noticable effects on the circuit operation.  This is because the resistor and capacitor from the compensated probe are already included in the circuit design so thier effects are already included in the normal operation.